Study on the Acoustic Matching Properties of High Frequency Medical Ultrasound Transducers Materials
Author:
Ethical statement:

Affiliation:

Funding:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
    摘要:

    医学超声成像的图像质量很大程度取决于超声换能器的特性。在换能器研制过程中,一项重要的任务是确定合适的匹配层材料及参数。为了满足高频超声换能器研制的需要,该文从理论及实验两方面研究了RTV615硅橡胶掺杂不同比例的氧化铝粉末后,所形成的复合材料的高频(20 MHz)声学特性参数,包括声速、声阻抗、声衰减系数等。结果表明,RTV615硅橡胶中掺杂氧化铝的体积分数为7.93%时,所形成的复合材料在20 MHz频率条件下声阻抗值提高52.94%,达到1.56×106 Pa·s/m,具有最佳的声阻抗匹配特性。因此,通过优化掺杂比例可有效改进硅橡胶材料的高频声学性能。

    Abstract:

    The quality of medical ultrasound imaging significantly depends on the performance of ultrasound transducer. During transducer fabricating, a key task is to determine appropriate materials and parameters of acoustic matching layers. To meet the requirements of the high frequency ultrasound transducer, in this paper the acoustic parameters of the composite of RTV615 silicone rubber mixed by alumina powder with different densities were investigated by theoretical models and experiments. The acoustic parameters include the sound speed, acoustic impendence and attenuation. The results show that the acoustic impendence of the 7.93% volume fraction of alumina doped composite is increased by 52.94%, up to 1.56×106 Pa·s/m at 20 MHz, which makes it more efficiently match with human tissue. It is a potential passive material for high frequency acoustic matching layer and transducer lens.

    参考文献
    相似文献
    引证文献
引用本文

刘鹏波,简小华,韩志乐,李章剑,崔崤峣.高频医用超声换能器材料声匹配特性研究[J].压电与声光,2016,38(3):450-453. LIU Pengbo, JIAN Xiaohua, HAN Zhile, LI Zhangjian, CUI Yaoyao. Study on the Acoustic Matching Properties of High Frequency Medical Ultrasound Transducers Materials[J]. PIEZOELECTRICS AND ACOUSTOOPTICS

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
历史
  • 收稿日期:
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2016-05-31
  • 出版日期: