Design of Ultrasonic Flow Velocity Profile Measurement System Based on FPGA
Author:
Ethical statement:

Affiliation:

Funding:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
    摘要:

    血流速度的检测对于临床医学具有重要意义。该文提出了一种基于现场可编程逻辑门阵列(FPGA)的脉冲多普勒血流速度剖面测量系统。该系统使用LiNbO3压电换能器作为探头,利用FPGA控制超声脉冲的收发;设计出基于希尔伯特变换的数字正交解调器取代传统混频器对回波信号进行解调,并在FPGA中实现快速傅里叶变换完成了频谱分析。经实验验证,系统能测出血管仿体内的速度剖面,测量误差小于7%。

    Abstract:

    Blood flow velocity detection is important for clinical medicine. An FPGA based system of ultrasonic pulsed wave Doppler was proposed to measure blood flow velocity profile. A LiNbO3 piezoelectric transducer was used as the probe to produce and receive ultrasonic pulses. The ultrasonic pulse recurrence frequency was controlled by FPGA. The conventional analog mixer was eliminated by the digital quadrature demodulator based on Hilbert transform for the echo signal demodulation. The spectrum analysis was realized in FPGA through the fast Fourier transform. The experiment verified that the proposed system can measure the velocity profile in blood vessel with the deviation of less than 7%.